High Performance LC
Fused silica capillary columns






                 
  Test-Chromatograms from Fused-Silica Nanocolumns          
  Using LC Nanocolumns easily improves the mass sensivity by a factor of up to
5000 compared with a 4.6 mm id column.
 
                   
  1.)
Dimension: 150 mm x 75 µm
Phase: Reprosil-Pur C18-AQ, 3 µm
Part.No.: R13.aq.s150.075
Batch-No.: 120305 86401

Eluent: ACN / water (70 /30)
Flow rate: 200 nl / min.
Pressure: 90 bars
Detection: UV 254 nm

Result:
Peak 1: Nitrobenzene , RT 2.96 min. N/m = 70 000
Peak 2: Biphenyl, RT 6.8 min.N/m = 120 000
   
                   
   
               
  2.)
Dimension: 200 mm x 75 µm
Phase: Reprosil-Pur 300 ODS-3, 3 µm
Part.No.: R33.93.s200.075
Batch-No.: 280905 16001

Eluent: ACN / water (70 /30)
Flow rate: 200 nl / min.
Pressure: 100 bars
Detection: UV 254 nm

Result:
Peak 1: Nitrobenzene, RT 4.65 min. N/m = 54 000
Peak 2: Biphenyl, RT 7.9 min. N/m = 95 000
 
           
   
               
  Increase of Mass Sensivity depending of inner column-diameter      
  Column-
id (mm)
Flow
(µl/min)
Increase of Sensivity        
  4.6 1000 1            
  2.0 200 4            
  1.0 40 21            
  0.5 10 85            
  0.3 3 235            
  0.1 0.5 2100            
  0.075 0.3 3760            
                   
                   
  Prices for Capillary Columns in Euro      
  Fused Silica capillary columns with 50 µm, 75 µm, 100µm, 150 µm  id  
  Stainless steel GLT-columns with 0.3mm0.4mm 0.5mm 0.75mm id  
 
               
  Prices for all kind of silica HPLC-phases:          
  For polymeric phases  (Poros, Hamilton):  + 50 Euro extra.)          
  Column-Length in mm:     40 125 150 200 250  
 
  Part-no.:     ..s04xxx ..s12xxx ..s15xxx ..s20xx ..s25xxx  
  Inner diamter 50 µm – 0.75 mm:     355,- 370,- 375,- 380,- 400,-  
 
               
  Bulk material on request                
 
               
 
  up