High Performance LC
ReproSil-Pur® Basic - the validated column


The perfect column with outstanding performance and stability:
These Basic C-18 phases are available with following particle sizes:
10µm --> for prep. HPLC
5 + 3µm --> for analytical HPLC
1,8µm --> for ultrafast




         
  Features:        
  - C18 monolayer bonded to spherical particles with 100 A° pores, 450 m2/g and a pore volumne
of 1.1 ml/g
 
  - Impurities for metals: Al, Ti, Fe, Zr:  all < 1 ppm
 
  - Typical plates / m between 80 and 100 000 for 5 µm
 
  - pH-stable from 1 up to 10 because of
 
  - double bound endcapping: unique stable
 
 
  Each batch is validated with following 6  quality tests:
 
  Test 1:        
  Inactivity of Silanols:        
  k (Pyridine) / k (Phenyl) Specification:  < 0.45      
  (eluent: MeOH / water, 7/3        
   
       
  Test 2:        
  Retentivity for Hydrophilic Compounds:      
  (column 150 x 4.6 mm, 1 ml flow / min.)      
  (eluent: 100 % water) Specification:  > 9 min.      
 
       
  Test 3:        
  Hydrogen-Bonding Capacity:        
  (eluent: MeOH / water, 8/2)        
  k (Coffeine) /  k (Phenol) Specification: 0.5 –0.6      
  Hydrophobicity:        
  k (n-Amylbenzene) / k (n-Butylbenzene) Specification:145 – 1.50      
  Shape-Selectivity:        
  k (Triphenylene) / k (o-Terphenyl) Specification: 1.20 – 1.30      
 
       
  Test 4:        
  Peak Symmetry for Basic Compounds:      
  (eluent: 20 mM K-Phosphate buffer pH 7 / ACN)      
  As (Propranol) Specification: <1.3 typical: 1.0 – 1.1
  As (Amitriptyline) Specification: < 1.6 typical: 1.1 – 1.3  
 
       
  Test 5:        
  Peak Symmetry for Acidic Compounds:    
  (eluent: 20 mM K-Phosphate buffer pH 7 / ACN)    
  k (p-Toluic Acid / k (Methyl Benzoate) Specification: < 0.45      
  As (p-Ethylbenzoic acid) Specification: < 1.4 typical: 0.9 – 1.1  
 
       
  Test 6:        
  Metal Sensivity:        
  (eluent: ACN / water, 1/1)        
  As (2,3 Dihydroxynaphthalene) / AS (2,7  Dihydroxynaphtalene) 1.5 typical: 0.9 – 1.1  
 
       
  Comparison of Inertsil ODS-3V and Reprosil-Pur Basic C18  
     
           
  Comparison of different C18 Phases
 
       
           
   
   